Product Description
Technical Specification
Accessories
Data Sheet

Elcometer 6910 Setaflash ‘Series 8’ Active-Cool (Corrosion Resisting)

The Elcometer 6910 Setaflash ‘Series 8’ Active-Cool (Corrosion Resisting) offers the same feature as the Active Cool Closed Cup Tester but comes with a corrosion resisting insert in the cup for testing corrosive samples.

It has been designed to carry out flash/no flash tests with a temperature range of 10°C to 130°C (50°F to 266°F) with no external cooling required. When connected to a suitable chilled water supply, the range is extended down to -20°C (-4°F).

 

There is also the Ramp Mode which increases the sample temperature at the rate of 2°C/min and automatically dips the ignitor every 1°C up to 100°C (every 2°C above 100°C), until either a flash is detected or the end of test temperature is reached.

 

• Small sample size, 2ml (0.067fl oz) or 4ml (0.135fl oz)
• Flash/No Flash & Ramp Modes
• Electric ignitor
• Automatic dipping and flash detection
• ActiveCool electronic Peltier cooling
• 64 test memory & RS232 interface
• Corrosion resisting insert

Elcometer 6910 Setaflash ‘Series 8’ Active-Cool (Corrosion Resisting)
Elcometer 6910 Setaflash ‘Series 8’ Active-Cool (Corrosion Resisting)

Technical Specification

Part Number Description Certificate 
K6910M022 Elcometer 6910 Setaflash ‘Series 8’ Active-Cool (Corrosion Resisting)
Temperature Range Air: 10°C to 130°C (50°F to 266°F)
Water Assisted: -20 to 130°C (-4 to 266°F)
Sample Size 2ml (0.067fl oz) or 4ml (0.135fl oz) according to method
Test Mode Rapid Equilibrium and Ramp
Test Duration Rapid Equilibrium Mode 1 minute below 100°C (212°F)      
2 minutes above 100°C (212°F)      
FAME 1 minute or user defined 1 to 99 minutes      
Test Duration Ramp Mode Typically 7 minutes
Cool Down Time 80°C to 20°C (176°F to 68°F) in 4 minutes
Cup Material Corrosion Resistant Steel Insert, Aluminium Cup
Dimensions 300 x 340 x 380mm (11.8 x 13.4 x 15.0”)
Weight 8kg (17.6lb)
Packing List Elcometer 6910 Setaflash ‘Series 8’ Active-Cool (Corrosion Resisting), 2ml (0.067fl oz) syringe, flash detector, mains power cable, programming plug, 2mm Allen key, 2.5mm Allen key, certified flash point material (50ml/1.7fl oz), certificate of calibration, warranty registration form and operating instructions


STANDARDS:
ASTM D3278, ASTM D3828, ASTM D7236, ASTM E502, BS 2000-523, CLP Regulations EC No 1272/2008, EPA 1020 A & B,

ISO 3679, ISO 3680, UN Class 3 Non-viscous Flammable Liquids

Standards in grey have been superseded but are still recognised in some industries.

 

Calibration Certificate supplied as standard.

Accessories

KT006910N009

Certified Flash Point Material (50ml/1.7floz)